all case studies
CO2 and Syngas Hydrogenation System
Catalyst Development
CO2-to-Chemicals
Parallel Reactor Screening
Syngas Hydrogenation
The system described here is a parallel reactor system with six reactors designed for catalyst testing in CO₂ hydrogenation and syngas hydrogenation reactions.
All reactors are operated at the same pressure and flow rate, while the temperature of each individual reactor can be controlled independently. This configuration enables highly reproducible comparative studies under defined reaction conditions.
Gas hourly space velocities (GHSV) between 1,000 and 50,000 h⁻¹ can be achieved with catalyst volumes of 5 mL.
Design Parameters
- Design temperature: 500 °C
- Design pressure: 150 bar
Reactor and Process Features
- 6 × double‑jacketed reactors
- Inner diameter: approx. 10 mm
- Isothermal zone: approx. 70 mm
- Parallel operation of all reactors
- Capillary‑based flow distribution
- Provision for an additional blank reactor
- Automatic sampling
- Connected to gas chromatograph / µ‑GC
Automation and Control
- Process control system: Siemens S7‑1500